Our WebCams and Virtual Instruments use Windows 2000 that can serve only 10 clients at one time. For this reason we must limit access. If you would like a temporary account and password to inspect our teaching methods, please send your request by e-mail to Prof. Bungay
C6H12O6 ---------> 2 C2H5OH + 2 CO2 (1)
(glucose) (ethanol) (carbon dioxide)
This equation shows that 0.511 g of ethanol and 0.489 g of CO2
are produced from each gram of glucose reacted. As some of the glucose
is used for the production and synthesis of secondary products and cell
components, the real stoichiometric yields are 90-95% of these values.
Accepting these approximate glucose conversion yields, it is possible to
follow the kinetics of a fermentation by measuring the mass of CO2
released. Another important aspect of alcoholic fermentation that employs
glucose as substrate and yeast as microorganism is the inhibition of glucose
consumption at high glucose concentrations. To avoid this, the fed-batch
fermentation is preferred. In this process, fermentation is started batchwise
with a low glucose concentration. When all the initial substrate is consumed,
a new addition restores the glucose concentration to just below the point
of where it produces inhibitory effects. It may be said that, by operating
in this way, the fed-batch fermentation is a sequence of batch fermentations
of increasing volumes. The apparatus for this experiment (shown in Figure
1) has a magnetically stirred 2-liter flask. Fresh fermentation medium
is supplied from another 2-liter flask via a computer controlled peristaltic
pump. Both the fermentation and medium flasks are on top-loading balances
equipped with serial computer interfaces to enable on-line calculation
of mass balances and productivity for the fermentation. The fermentation
flask is vented through a meter to provide an independent measure of the
CO2 released during ethanol production. A Macintosh computer
controls the pump and collects data.

We are going to rely on massive inoculation with Baker's yeast
and will circumvent sterilization. Usually, the feed reservoir must be
protected from contamination. As medium leaves, air enters. This air must
pass through dense fibrous material or a membrane that will filter out
microorganisms. The filter itself is sterilized as the reservoir and its
connections are autoclaved. Organisms in the fermenter can grow back through
the tubing. Yeast are big and slow and might not reach the reservoir in
24 hours. Bacteria however can grow back through the tubing in a few hours.
An air break device prevents this. Again, since we are running our experiment
for only 24 hours, we will not have to install an air break or a steril
filter. Each group will conduct one experiment. You may devise an original
experiment of your own. You can earn bonus points if you help the next
group to run the experiment. This may be advice before the laboratory session
or you may find a few moments to work with them during the session. You
only earn points if they include a brief acknowledgment when they write
their report.
Answer the following questions:
Discuss the answers with your TA.
1) Start up the computer: Push the upper right button on the key board. Let the machine power up completely before executing the next step.
2) Start up the software: Double click on the "Feedback of Balance Reading.vi" icon. Let the software start up completely before going to the next step. This is important to start up the serial communication properly.
3) Power up the remaining equipment: Push the switch on the panel. The pump has to show "P?1" The balances go through a self-test routine after the calibration bar is pressed. The indicator for temperature and gas flow rate will become illuminated. Open the valves upstream and downstream of the gas flow meter.
4) Install fresh tubing: This will be done by the TA. You must discuss with your TA whether fresh tubing should be installed. You can find a part list for number and length of the required tubing at the experiment. Make sure the tubing is connected exactly as it was before. The connections to the vessels must be horizontal loops. Vertical loops will add weight to the balance reading because the previously fresh tubing relaxes during the course of the experiment. If in doubt, you may copy the layout from the other setup. The pump must have a size 16 tubing.
5) Prepare the medium: You will need 2 liters of the following medium in Table 1.
TABLE 1 Medium composition (per liter of medium)
| KH2PO4 | 5 g |
| (NH4)2SO4 | 2 g |
| Carbon source (glucose) | 50 g |
| MgSO4.7H2O | 0.4 g |
| Yeast extract | 1 g |
| Deionized water | to one liter mark |
The amounts mentioned above will be good for 1 liter of medium. Deionized water can be found on the 3rd floor of Ricketts Building. The pH is adjusted to 5.5 with H3PO4. It is good practice to record the amounts actually added into your lab note book. Remove the feed flask from the balance! Make sure the rubber stopper properly seals the feed flask. Place the feed vessel back on the PM4000 balance.
6) Inoculation: Add 3g of freeze-dried yeast into the reactor. Remove the reactor from the balance! Be sure the rubber stopper properly seals the reaction vessel. Place the reactor back on the PM6100 balance. Center the stirrer bar on the magnetic stirrer unit.
7) Align the vessels: The vessels have to be in the center of the weighing pan (not the blue spill protection). The tubing leading to the vessels should not be nicked or under stress. Level the balances. A spirit level can be found at the back of the balances. Zero the balances.
8) Zero the gas flow meter: Pushing the [VIEW] buttom on the totalizer key board lets you select different readings on the display: r = gas evolution rate, P = peak point, u = valley point, no indicator = total gas volume. Pressing the [ENTER] key while the total gas volume is displayed will zero it. Only zeroing the total gas volume will influence the data send to the computer. Do not zero the total gas volume once the program is running. This will cause an error.
9) Enter the parameters into the program: Use a filename like: MyDisk:MyFile.dat. Do not save data to the hard disk since the device would suffer from constantly being accessed. It is cheaper to replace a floppy disk drive. Suggestion: Do not go above 40% nor below 10% of the glucose concentration in the feed for the setpoint. It is recommended using 0.025 g/ ml as setpoint for glucose. You have only one shot at this. Make sure you set all the parameters correctly. It is also recommended selecting the gas evolution rate for the totalizer reading.
10) Zero balances: There must be clearance between the balance body and the blue spill protection. Again, check zero of the totalizer for the gas flow meter and the balances. You must not touch anything once the program is running!
11) Start the program: Click with the mouse on the arrow button in the upper left corner of the window. The running program will initialize the pump and change its display to "P01''. Within 10 minutes, the pump will feed about 300 ml of medium into the reactor.
12) Start the stirrer: Wait until the pump stopped. Do not exceed 1/2 of the maximum stirrer speed. You are in big trouble, once the stirrer spins out of control. Do not touch the electric connection of the stirrer since this could change the zero point of the balance.
13) Operation: Observe the reactor for one hour. The pump must come on for 10 minutes after two hours, and add some more medium to the reactor. The flow meter should start recording a gas flow.
14) Before leaving the lab: Setup a time with your TA to return the next day. The next day
15) Initial observations: The feed flask should now be empty. The totalizer should show a certain total gas volume.
16) Record data: Record the elapsed time, the total gas volume, the current gas flow rate, the temperature reading and the readings of the balances in your lab note book.
17) Stop the program: Click the red stop button on the screen. Print the front panel by pressing the [Apple key] and the [P] simultaneously.
18) Close the application program: Do not save the program.
19) Check Data: Double click on the disk icon and verify that the collected data are really on the disk. The data are saved in a Macintosh spreadsheet format. The individual columns are: Column# = time in [min], PM4000 [g], PM6100 [g], Pump [ml/min], Totalizer [ml/min, or ml; as selected], Temperature [C], Go (Glucose conc. at begin of batch), Gt (estimated glucose conc. in reactor), Mco2 (Mass of CO2 expected to evolve during current batch). The installed software allows saving the data to a MS-DOS disk or a Macintosh disk.
20) Shut down: Clean all equipment and power everything down.
Close the valves upstream and downstream of the gas flow meter. It can
damage the flow meter if you forget to close the valves! Leave the old
tubing in place.
FLUSH THE TUBING WITH TAP WATER TO REMOVE ORGANISMS AND CRUD.
Points to consider
Is this a real fed-batch?
Compare your experimental setup with the theory of fed-batch fermentation.
Try to explain your data. You may try to simulate your experiment.
What is the state of the yeast before inoculation? Is it alive, is it dead?
What is the difference between the freeze-dried yeast and yeast extract?
The software will save the following values: Column# = elapsed
time [min], PM4000 [g], PM6100 [g], Pump [ml/min], Totalizer [ml/min, or
ml; as selected], Temperature [C], Go (Glucose conc. at begin of batch),
Gt (estimated glucose conc. in reactor), Mco2 (Mass of CO2
expected
to evolve during current batch). Data points are saved at intervals of
1 minute.
6. GRADING POLICY
The scores for presentations are covered in the web page about the interim. In most cases, all members of a team get the same grade. However, a team member who has contributed much better or much worse than the others may get a different grade.
INCLUDING MATERIAL FROM AN OLD REPORT MAY CAUSE YOUR GRADE TO BE
"F".
Bailey, H.R. and Ollis, D.F. (1986) "Biochemical Engineering Fundamentals". McGraw-Hill, Inc.
Bungay, H.R. (1992). "Basic Biochemical Engineering". BiLine Associates, Troy NY.
Video. "The Fermentation Lab". Department of Chemical Engineering.
Rensselaer Polytechnic Institute, Troy NY.
Rate the following items, 10 points being the maximum, X meaning
no opinion:
| Quality of the experiment compared to other senior lab experiments | (0 to 10) |
| Quality of the experiment on an absolute basis | - |
| Enjoyment of experiment | - |
| New knowledge gained | - |
| Clarity of instructions | - |
| User-friendliness of software | - |
| Quality of instruction | - |
| Other sources of information (e.g., reserve literature) | - |
Answer the following questions on a separate sheet of paper:
1) What are positive aspects of the lab handout?
2) How could the lab handout be improved?
What would you like to be added to it?
What should be removed?
What needs to be corrected?
3) What are positive aspects of the experimental setup?
4) What are negative features of the setup?
Is anything broken?
How user-friendly is the setup?
Do you think there is a fundamental bug in the setup?
5) How could the hardware be improved?
6) Was your TA competent to answer your questions?
8) Did you understand how to obtain a good grade?
9) Please feel free to add anything, negative or positive.